High-Accuracy Digital Contact SensorGT2 series

GT2-P12KL

Stylus type High-precision Sensor head (Low measuring force type)

SPECIFICATIONS

Select Language

Model

GT2-P12KL*1

Detection system

Scale Shot System II, absolute (no tracking errors) type

Measuring range

12 mm 0.47"

Resolution

0.1 μm 0.004 Mil

Indicated accuracy

1 μm 0.04 Mil (p-p)*2

Measuring force

Downward mounting

0.2 N*3

Side mounting

0.15 N*3

Upward mounting

0.1 N*3

Sampling cycle

4 ms

Mechanical response

4 Hz*2

Operation indicator light

2-color LED (red, green)

Environmental
resistance

Enclosure rating

Ambient temperature

-10 to +55°C 14 to 131°F (No freezing)

Relative humidity

35 to 85% RH (No condensation)

Vibration

10 to 55 Hz Double amplitude 1.5 mm 0.06" in the X, Y, Z axis directions respectively, 2 hours

Environmental resistance

Impact resistance

1000 m/s2(IEC60068-2-27)

Materials

Main body

Main body case: SUS 303, Status indicator: PET, Sensor head-relay connector cable: PUR, Relay connector: PBT

Dust boot

-

Contact

SUS304, cemented tungsten carbide*4

Sensor head cable

Optional (connect to relay connector)

Weight

Approx 35 g (excluding cable)*5

*1 Note: You may not be able to connect the sensor head to the amplifier unit depending on when the amplifier unit was purchased.For details, contact your local sales office.
*2 Value when the ambient temperature is 20°C 68°F.
*3 Value at center of measuring range. Please note that the measurement force changes depending on the installation state of the dust boots.
*4 The contact is included with the sensor.
*5 Including the relay connector.

Contact Sensor Tips