Ultra-High Speed In-line ProfilometerLJ-V series

LJ-V7001

Controller English Manual

SPECIFICATIONS

Select Language

Model

LJ-V7001

Number of connectable sensors

Max. 2 units

Display

Minimum
displayable unit

0.1 µm, 0.00001 mm2, 0.01°

Maximum display range

±9999.99 mm 393.7", ±9999.99 mm2

Input terminal block

Laser remote interlock input

No-voltage input*1

Input
terminal block

Encoder input

NPN open collector output/PNP open collector output (5V / 12V / 24V) line driver output

Input terminal block

Trigger input

No-voltage input*1

Timing 1,2 input

Auto-zero 1,2 input

Input
terminal block

Reset 1, 2 input

Input terminal
block

Start measurement/stop input

Start storage/stop input

Clear memory input

Laser OFF input

Program switch input

No-voltage input x 4 inputs*1

Output terminal block

Analog voltage output

±10 V x 2 outputs, Output impedance: 100 Ω

Output
terminal block

OUT comparator output

NPN open collector output x 12 outputs*2
(Can freely assign 16 OUTs x 3 stage judgment results)

Strobe output

NPN open collector output*2

Disable trigger output

Memory FULL output

Ready output

Output terminal block

Error output

NPN open collector output (N.C.)*2

PC interface

Ethernet

1000BASE-T/100BASE-TX

USB

USB 2.0 high speed compliant (USB 1.1 Full-SPEED compatible)

RS232C

Measurement data output and control I/O (Can select a baud rate of up to 115,200 bits/s)

Rating

Power voltage

Includes 24 VDC ±10% ripple (P-P)

Current consumption.

1.3 A or less when connected to 1 head/ 1.9 A or less when connected to 2 heads

Environmental resistance

Operating temperature range

0 to +50°C 32 to 122°F

Operating ambient humidity

20 to 85% RH (No condensation)

Weight

Approx 1500 g

*1 The rating for non-voltage input is up to 1 V for ON voltage and up to 0.6 mA for OFF current
*2 The rating for NPN open-collector output is up to 50 mA (40 V or less), residual voltage of up to 1 V

Tips of Measurement