Laser scan micrometer

LS-3000 series

Select Language

Model

Scanning head

LS-3034

LS-3033SO(8073)

LS-3032

LS-3033

LS-3036

LS-3060

Standard controller/2-head controller

LS-3100

LS-3100W/LS-3100DW

LS-3100

Type

Fine measuring

Standard

Wide range

Scanning head type

Single-body

Separate

Single-body

Separate

Single-body

Separate

Measuring range

0.08 to 30 mm 0.003" to 1.18"*1

0.3 to 30 mm
0.01" to 1.18"

0.5 to 30 mm
0.02" to 1.18"

0.3 to 30 mm
0.01" to 1.18"

0.8 to 60 mm
0.03" to 2.36"

Measurement accuracy

±2 µm max.*2 ±0.000079"

±3 µm max.*5
±0.000119"

Minimum target width

0.08 mm 0.003"

0.3 mm 0.01"

0.5 mm 0.02"

0.3 mm 0.01"

0.8 mm 0.03"

Transmitter/receiver distance

60 mm 2.36"*3

120 mm 4.72"
(Fixed)*3

60 mm 2.36"*3

120 ±40 mm
4.72" ±1.57"*3

60 mm 2.36"*3

200 ±50 mm 7.87" ±1.97"*3

Repeatability (number of averaging measurements: 512)

±0.3 µm ±0.000012"

±0.5 µm ±0.000020"

Minimum displayable unit

0.1μm

Laser scan rate

400 scans/s

Laser scan velocity

63 m 206.7'/s

126 m 413.4'/s

Laser scan range

Approx. 33 mm 1.30"

Approx. 65 mm 2.56"

Light source

Semiconductor laser, wavelength: 780 nm, output:55μ, Class 1 laser product (JIS C6802)

Semiconductor laser, wavelength: 670 nm, output:35μ, Class 1 laser product (JIS C6802)

Semiconductor laser, wavelength: 670 nm, output:25μ, Class 1 laser product (JIS C6802)

Display

Minimum display unit

0.1 µm

Terminal block output

Control output

Analog output voltage

±6 V, Output impedance: 0 Ω

Control output

SPST-NO Relay contact 250 VAC 2 A (resistive load)

Control input

Hold timing input

Non-voltage input (contact/solid state)

Laser remote input

Program switching input (program No. 1 to 5)

36-pin connector I/O

Hold timing input

TTL, level, negative logic

Auto-zero set/reset input

Program switching input (program No. 1 to 5)

Control output

7-level judgment output

Maximum NPN open-collector output: 40 mA (30 V or less)

3-step limits output

Interface

RS-232C

Measurement data output and control output (baud rate: selectable between 75 and 19,200 bps)

GP-IB

Measurement data output and control output*4

BCD

Measured data output (8 digits)*4

Rating

Power voltage

85 to 264 VAC, 50/60 Hz

Power consumption

40 VA max.

Environmental resistance

Enclosure rating

IP60

Ambient
temperature

Scanning head

0 to 45°C 32 to 113°F

0 to +40℃ 32°F to 104°F

Controller

0 to 45°C (32 to 113°F)

Environmental resistance

Operating ambient humidity

35 to 85%, No condensation

Weight

Scanning head

0.8 kg*5

T: 0.5 kg
R: 0.4 kg*5

0.8 kg*5

T: 0.5 kg
R: 0.4 kg*5

0.8 kg*5

Approx. 2.5 kg
(base included)*5

Controller

Approx. 4.6 kg

*1 When the outer diameter is measured.
*2 When a rod 10 mm 0.39" in diameter is measured in a 2 x 20 mm 0.08" x 0.79", 4 x 20 mm 0.16" x 0.79", or 10 x 20 mm 0.39" x 0.79" measuring area.
*3 T' indicates a transmitter, ‘R' indicates a receiver.
*4 Supports options