High-speed, High-accuracy Digital Micrometer

LS-7000 series

Select Language

Model

LS-7601

LS-7001

Type

High-performance

Standard

No. of connectable measuring heads

2 (All types)

2 (LS-7070/LS-7030/LS-7010 only)

Display

Measurement display

TFT 5.5-inch LCD display,
Backlight: CCFL (Average life: 40,000 hours)

Main display: 7-segment red LED
(Character height: 20.3 mm 0.80")
Sub-display: 7-segment red LED
(Character height: 9.9 mm 0.39") x 3

Minimum display unit

0.01 µm to 100 µm (7-level selectable)

Display range

±99.99999 to±9,999.9 mm ±3.937" to ±393.7" (linked to the minimum display unit, mm/ µm selectable)

Measurement position monitor

Monitor image (When the measuring head
with the monitor function is connected.)

7-level display with a red LED

Tolerance check output display

5-level LCD indicator

Green LED (GO), Red LED x 2 (HI, LO)

Terminal block

Alarm output

NPN open-collector output (N.C.)*1

5-level judgment output
judgment ready output
strobe output

NPN open-collector output for OUT1*1

Comparator ready output

Program selection input

Non-voltage input x 4 inputs*1

Strobe output

NPN open-collector output for OUT1*1

Statistical processing input

Non-voltage input for OUT1*1

*1

Timing input
reset input
auto-zero input

Non-voltage input for OUT1*1

Analog output

±10 V x 2 outputs*1

Reset input

Non-voltage input for OUT1*1

Auto-zero input

Connector I/O

SUB mode

Statistical processing output

NPN open-collector output x 2 outputs*2

*2

Function output

Selectable from focus, area check, and difference check, NPN open-collector output x 2 outputs*2

5-level judgment output judgment ready output strobe output

NPN open-collector output for OUT2*2

Comparator ready output

Strobe output

BCD mode

BCD output

Measurement data output (Sign + 7 digits), OUT1/OUT2 selectable, NPN open-collector output*2

BCD selection output

NPN open-collector output*2

BCD selection input

Non-voltage input*2

Statistical processing input

Non-voltage input for OUT2

Timing input reset input auto-zero input

Non-voltage input for OUT2

Reset input

Auto-zero input

RS-232C interface

Measurement data output and control I/O, printer (Baud rate can be selected up to 115.200 bps)

Video output

Conforming to the NTSC system (RCA connector)

Major functions

Simultaneous measurement, area designation, calculation, averaging, calibration, 16-program memory, measurement, auto-zero, print-out, abnormal value elimination, transparent object measurement, measurement point display, group comparison, unacceptable va

Simultaneous measurement, area designation, calculation, averaging, calibration, 16-program memory, measurement, auto-zero, print-out, abnormal value elimination, transparent object measurement, mutual interference prevention, etc.

Rating

Power voltage

24 VDC ±10%*3

Power consumption

1.2 A max.*3

0.7 A max.*3

Environmental resistance

Enclosure rating

IP64 (Panel surface only)

Operating temperature range

0 to +40℃ 32°F to 104°F

Operating ambient humidity

35 to 85% RH (No condensation)

Weight

Approx 1010 g

Approx 820 g

*1 The rating of the NPN open-collector inside the terminal block is: 100 mA max. (40 V max.), with a residual voltage of 0.5 V max. The rating of the NPN open-collector inside the connector I/O is: 30 mA
max. (30 V max.), with a residual voltage of 0.5 V max. The rating of non-voltage input is: ON voltage of 1 V max., OFF current of 0.6 mA max.
*2 Either SUB mode or BCD mode can be selected.
*3 AC power supply can be used when the LS-S11 (AC power supply stand) is connected.

Model

LS-7070M

LS-7070

LS-7030M

LS-7030

LS-7010M

LS-7010

Type

Large-diameter type

Wide measurement range

Super fine measurement

Monitor camera

Provided

Not provided

Provided

Not provided

Provided

Not provided

Classification

Measurement part with monitor function

General-purpose measurement part

Measurement part with monitor function

General-purpose measurement part

Measurement part with monitor function

General-purpose measurement part

Measuring range

0.5 to 65 mm 0.02" to 2.56"

0.3 to 30 mm 0.01" to 1.18"

0.04 to 6 mm 0.002" to 0.24"

Minimum detectable object

0.5 mm0.02"

0.5 mm 0.02"

0.3 mm 0.01"

0.04 mm 0.002"

Transmitter/receiver distance

250 ±50 mm 9.84" ±1.97"

160 ±40 mm 6.30" ±1.57"

60 mm 2.36"

Light source

GaN green LED

CCD scanning range

Approx. 69 mm 2.72"

Approx. 33 mm 1.30"

Approx. 7 mm 0.28"

Measurement accuracy

±3 µm ±0.000119"l *1

±3 µm ±0.000119" *1

±2 µm ±0.000079" *5

±0.5 µm ±0.000020" *7

Repeatability

±0.2 µm ±0.000008" *2

±0.15 µm ±0.000006" *6

±0.06 µm *8

No. of samples

2,400 samples/sec.*3

Monitor function

Provided

Not provided

Provided

Not provided

Provided

Not provided

Environmental resistance

Enclosure rating

IP64*4

Operating temperature range

0 to +50°C 32 to 122°F

Operating ambient humidity

35 to 85% RH (No condensation)

Weight

Transmitter: Approx.540 g,
Receiver: Approx. 770 g,
Base: Approx. 660 g

Transmitter: Approx.540 g,
Receiver: Approx. 730 g,
Base: Approx. 660 g

Transmitter: Approx.420 g,
Receiver: Approx. 570 g,
Base: Approx. 430 g

Transmitter: Approx.420 g,
Receiver: Approx. 470 g,
Base: Approx. 430 g

Transmitter: Approx.140 g,
Receiver: Approx. 380 g,
Base: Approx. 220 g

Transmitter: Approx.140 g,
Receiver: Approx. 340 g,
Base: Approx. 220 g

*1 The error when a moving rod 20 mm 0.79" in diameter is measured in the measuring area of 20 x 40 mm 0.39" x 0.79".
*2 The value of±2σ when the outer diameter of a rod 20 mm 0.79" in diameter is measured at the center of the measuring area while the number of averaging measurements is set to 512.
*3 1200 samples/sec. when the mutual interference prevention function is used.
*4 The connector section is excluded.
*5 The error when a moving rod 10 mm 0.39" in diameter is measured in the measuring area of 10 x 20 mm 0.39" x 0.79".
*6 The value of±2σ when the outer diameter of a rod 10 mm 0.39" in diameter is measured at the center of the measuring area while the number of averaging measurements is set to 512.
*7 The error when a moving rod 1.0 mm 0.04" in diameter is measured in the measuring area of 2 x 4 mm 0.08" x 0.16".
*8 The value of±2σ when the outer diameter of a rod 1.0 mm 0.04" in diameter is measured at the center of the measuring area while the number of averaging measurements is set to 512.

Model

LS-H1W

Supported operating systems

Windows Vista, Windows XP Professional Edition/Home Edition,
Windows 2000 Professional*1

Required free space on hard disk

10 MB or more

Required display

VGA (800 x 600 pixels) or more, 256 colors or more

CPU

Pentium III 400 MHz or higher

Memory capacity

64 MB or more

PC interface

RS-232C (serial port) or USB to
RS-232C converter (OP-51643) interface required

Remarks

Excel 2003/2002/2000*1

*1 Windows Vista, Windows XP Professional Edition/Home Edition, Windows 2000 Professional, Excel 2003/2002/2000 are registered
trademarks of Microsoft Corporation in the U.S.A.

Model

LS-S11

Applicable controller

LS-7001/LS-7601

Rating

Power voltage

100 to 240 VAC ±10%, 50/60 Hz

Power consumption

110 VA max.

Environmental resistance

Operating temperature range

0 to +40℃ 32°F to 104°F

Operating ambient humidity

35 to 85% RH (No condensation)

Weight

Approx 1.7 kg

Model

LS-C3A

LS-C10A

LS-C30A

LS-C3AM

LS-C10AM

LS-C30AM

Product Name

Head-Controller Cable 3-m

Head-Controller Cable 10-m

Head-Controller Cable 30-m

Head-Controller Camera Cable 3-m

Head-Controller Camera Cable 10-m

Head-Controller Camera Cable 30-m

Overview

Cable between the controller and measuring head

Camera cable between the controller and measuring head

Applicable product

LS-7000 series

Number included

1

Length

3m

10m

30m

3m

10m

30m

End of cable 1

Straight

Tips of Measurement

  • General Catalog Request
  • Trade Shows and Exhibitions
  • eNews Subscribe

Measurement Sensors