High-speed optical micrometerLS-9000 series

LS-9501

Controller

SPECIFICATIONS

Select Language

Model

LS-9501*1*2

No. of connectable sensor heads

2

Head compatibility

Yes

Interface

Ethernet

1000BASE-T/100BASE-TX

USB

USB 2.0 HI-SPEED supported (USB 1.1 Full-SPEED compatible)

RS-232C

Measured value output, control I/O, setting change, baud rate can be selected up to 115,200 bps

Display and settings panel interface

LS-D1000 Max. four heads connectable

Display

Minimum display unit

0.01 µm

Display range

±99999.99 µm to ±9999.9 mm

LED display

POWER ON indicator, ERROR indicator

Input terminal block

Encoder input

NPN open collector output/PNP open collector output (5V / 12V / 24V) line driver output

Synchronous 1, 2 input

Non-voltage input

Auto-zero 1, 2 input

Reset 1, 2 input

Storage trigger input

Storage enable input

Storage data clear input

Statistics 1, 2 input

Statistics clear 1, 2 input

Program selection input

Non-voltage input × 4 inputs

Output terminal

Analog voltage output

±10 V × 2 outputs, output impedance100 Ω

Analog current output

4 to 20 mA × 2 outputs, compatible load max. 350 Ω

Universal output

NPN open-collector output x 10 outputs
Measured value and tolerance judgment output,
status output allocatable

Status 1, 2 output

NPN open-collector output

Total judgment output

Memory FULL output

Strobe 1, 2 output

Error output

NPN open-collector output (N.C.)

Rating

Power voltage

24 VDC ±10%, including ripple (P-P)

Current consumption.

When LS-HA100 not used: 1.0 A max. when 1 head connected; 1.3 A max. when 2 heads connected
When LS-HA100 in use: 1.8 A max. when 3 heads connected; 2.1 A max. when 4 heads connected*3

Environmental resistance

Operating temperature range

When LS-HA100 not used: 0 to +50°C 32 to 122°F
When LS-HA100 in use: 0 to +45°C 32 to 113°F

Operating ambient humidity

20 to 85% RH (no condensation)

Weight

Approx 1,500 g

*1 NPN open-collector output rating: 50 mA max. (40 V max.), residual voltage of 1 V max.
*2 Non-voltage input rating: ON voltage of 1 V max., OFF current of 0.6 mA max.
*3 Add the current consumption values for all units when connecting the display settings panel and expansion units.