High-speed optical micrometer

LS-9000 series

Select Language

Controller

Model

LS-9501*1*2

LS-9501P*4*5

No. of connectable sensor heads

2

Head compatibility

Yes

Interface

Ethernet

1000BASE-T/100BASE-TX

USB

USB 2.0 HI-SPEED supported (USB 1.1 Full-SPEED compatible)

RS-232C

Measured value output, control I/O, setting change, baud rate can be selected up to 115,200 bps

Display and settings panel interface

LS-D1000 Max. four heads connectable

Display

Minimum display unit

0.01 µm

Display range

±99999.99 µm to ±9999.9 mm

LED display

POWER ON indicator, ERROR indicator

Input terminal block

Encoder input

NPN open collector output/PNP open collector output (5V / 12V / 24V) line driver output

Synchronous 1, 2 input

Non-voltage input

Voltage input

Auto-zero 1, 2 input

Reset 1, 2 input

Storage trigger input

Storage enable input

Storage data clear input

Statistics 1, 2 input

Statistics clear 1, 2 input

Program selection input

Non-voltage input × 4 inputs

Voltage input × 4 inputs

Output terminal

Analog voltage output

±10 V × 2 outputs, output impedance100 Ω

Analog current output

4 to 20 mA × 2 outputs, compatible load max. 350 Ω

Universal output

NPN open-collector output x 10 outputs
Measured value and tolerance judgment output,
status output allocatable

PNP open-collector output x 10 outputs
Measured value and tolerance judgment output,
status output allocatable

Status 1, 2 output

NPN open-collector output

PNP open-collector output

Total judgment output

Memory FULL output

Strobe 1, 2 output

Error output

NPN open-collector output (N.C.)

PNP open-collector output (N.C.)

Rating

Power voltage

24 VDC ±10%, including ripple (P-P)

Current consumption.

When LS-HA100 not used: 1.0 A max. when 1 head connected; 1.3 A max. when 2 heads connected
When LS-HA100 in use: 1.8 A max. when 3 heads connected; 2.1 A max. when 4 heads connected*3

Environmental resistance

Operating temperature range

When LS-HA100 not used: 0 to +50°C 32 to 122°F
When LS-HA100 in use: 0 to +45°C 32 to 113°F

Operating ambient humidity

20 to 85% RH (no condensation)

Weight

Approx 1,500 g

*1 NPN open-collector output rating: 50 mA max. (40 V max.), residual voltage of 1 V max.
*2 Non-voltage input rating: ON voltage of 1 V max., OFF current of 0.6 mA max.
*3 Add the current consumption values for all units when connecting the display settings panel and expansion units.
*4 PNP open-collector output rating: 50 mA max. (30 V max.), residual voltage of 1 V max.
*5 Voltage input rating: Input max. voltage 26.4 V, min. ON voltage 10.8 V, OFF current 0.6 mA max.

Head

Model

LS-9006M*1

LS-9006*1

LS-9030M*1

LS-9030*1

Type

with monitor camera

without monitor camera

with monitor camera

without monitor camera

Measurement range

0.04 mm (0.01 mm) to 6 mm 0.002" (0.0004") to 0.24"

0.3 mm (0.08 mm) to 30 mm 0.01" (0.003") to 1.18"

Smallest detectable object

0.04 mm (0.01 mm) 0.002" (0.0004")

0.3 mm (0.08 mm) 0.01" (0.003")

Transmitter/receiver distance

60 ±5 mm 2.36" ±0.2"

160 ±40 mm 6.3" ±1.57"

Repeatability

±0.03 µm*2

±0.1 µm 0.000004"*6

Measurement accuracy

±0.5 µm 0.000020"*3

±2 µm 0.000079"*7

Sampling cycle

16,000 samples/sec.*4

Transmitter/receiver direction and position detection

Detection area

4 × 5 mm 0.16" x 0.2"

20 × 24 mm 0.79" x 0.94"

Smallest detectable object

0.04 mm 0.002"

0.3 mm 0.01"

Repeatability

±0.02 mm 0.0008"*5

±0.2 mm 0.01"*8

Sampling cycle

4,000 samples/sec.

Light source

InGaN green LED

Monitor camera

Provided

Not provided

Provided

Not provided

Rating

Current consumption.

with monitor camera

without monitor camera

with monitor camera

without monitor camera

Environmental resistance

Operating temperature range

0 to +50°C 32 to 122°F

Operating ambient humidity

20 to 85% RH (no condensation)

Operating ambientluminance

Incandescent lamp/fluorescent lamp 3,000 lux or lower

Vibration

10 to 55 Hz, double amplitude 1.5 mm 0.06", 2 hours in each direction (X,Y, and Z)

Shock resistance

15G/6 ms

Enclosure rating

IP67 (including connector)

Material

Aluminum

Weight

Transmitter: Approx 130 g
Receiver: Approx 300 g
Base: Approx 180 g

Transmitter: Approx 130 g
Receiver: Approx 280 g
Base: Approx 180 g

Transmitter: Approx 440 g
Receiver: Approx 500 g
Base: Approx 430 g

Transmitter: Approx 440 g
Receiver: Approx 440 g
Base: Approx 430 g

*1 The values in brackets are measured in ultra-thin mode.
*2 A ±2σ margin of error when measuring a ø1.0 mm ø0.04" rod in the center of the measurement area using outer diameter mode with the average measurement number set as 2048 times.
*3 Margin of error when a moving ø1.0 mm ø0.04" rod is measured in the 2 mm × 4 mm 0.08" × 0.16" measurement area using outer diameter mode.
*4 The sampling cycle is changed by the number of OUT set, and by the use of the mutual interference prevention function.
*5 The value of ±2 when the position of a rod ø1.0 mm 0.004" is measured at the center of the measuring area while the number of averaging measurements is set to 512.
*6 A ±2σ margin of error when measuring a ø10 mm ø0.39" rod in the center of the measurement area using outer diameter mode with the average measurement number set as 2048 times.
*7 Margin of error when a moving ø10 mm ø0.39" rod is measured in the 10 mm × 20 mm 0.39" × 0.79" measurement area using outer diameter mode.
*8 The value of ±2 when the outer diameter of a rod ø10 mm 0.39" is measured at the center of the measuring area while the number of averaging measurements is set to 512.

Head expansion unit

Model

LS-HA100

No. of connectable sensor heads

2

Head compatibility

Yes

Display

LED display

POWER ON indicator, head status indicator

Output terminal

Analog voltage output

±10 V × 2 outputs, output impedance100 Ω

Analog current output

4 to 20 mA × 2 outputs Compatible load max. 350 Ω

Power supply

Supplied from the controller

Environmental resistance

Operating temperature range

0 to +45°C 32 to 113°F

Operating ambient humidity

20 to 85% RH (no condensation)

Weight

Approx 600 g

OS environment for using the LS-H2 (LS-Navigator 2) Setting Support Software

Model

LS-H2

Supported operating systems

·Windows 7*1
·Windows Vista*2
·Windows XP (SP2 or later)*3

Languages

Japanese, English, German, Simplified Chinese, Traditional Chinese

Required free space on hard disk

10 GB or more

Required display

XGA (1024 × 768 pixels) or more, 256 colors or more

CPU

Core 2 Duo 2 GHz or more

Memory capacity

2 GB or more

L2 cache memory

2 MB or more

PC interface

Ethernet 1000BASE-T/100BASE-TX*4

*1 Each edition of Home Premium, Professional, and Ultimate are supported.
*2 Each edition of Ultimate, Business, Home Premium, and Home Basic are supported.
*3 Home Edition and Professional Edition are supported.
*4 Connection to a LAN and connection via a router is not covered by the warranty.

Display and settings panel

Model

LS-D1000

Type

0.19 A or less

Display interface

Measured value display

Measured value display: 2 colors, 8 digits, 16 segments
OUT number display: Monochrome, 2 digits, 7 segments
Tolerance judgment display: HH, HI, GO, LO, LL. Monochrome
Control status display: TIM, ZERO indicator. Monochrome

Program number display

Monochrome, 2 digits, 7 segments

Position monitor display

1D display: 2 colors, 32 levels
2D display: Monochrome, 7 x 7 matrix display

Display update cycle

5 times/sec.

Operation input interface

Numeric keypad, function key, lock key
timing input key, zero input key, reset input key, escape key,
arrow keys (4)

Display and settings panel connection port

2

Power supply

Supplied from the controller

Rating

Current consumption.

0.19 A or less

Environmental resistance

Operating temperature range

0 to +50°C 32 to 122°F

Operating ambient humidity

20 to 85% RH (no condensation)

Enclosure rating

IP65 (When panel attached, front surface only)

Weight

Approx 400 g

Transmitter-receiver cable

Model

OP-87686

OP-87687

Cable min. bending R

10 mm 0.39"

Enclosure rating

IP67 (IEC60529)*1

Length

1 m 3.3'

3 m 9.8'

Material

PVC

Weight

Approx. 60 g

Approx. 130 g

*1 Only applies when connected to both the transmitter and receiver.

Dedicated cable for display panel

Model

OP-87602

OP-87603

OP-87604

OP-87605

Cable min. bending R

24 mm 0.94"

Enclosure rating

None

Length

2 m 6.6'

5 m 16.4'

10 m 32.8'

20 m 65.6'

Material

PVC

Weight

Approx. 150 g

Approx. 350 g

Approx. 700 g

Approx. 1,400 g

40 m extension connection unit

Model

CB-BR01

Type

Extend communication between heads and controllers

Cable min. bending R

22 mm 0.87"

Power supply

Supplied from the controller

Environmental resistance

Operating temperature range

0 to +50°C 32 to 122°F

Operating ambient humidity

20 to 85% RH (no condensation)

Enclosure rating

None

Weight

Approx 150 g (including cable)

BCD output unit

Model

CB-BD100*1*2*3

Display

LED display

POWER-ON LED

Input terminal

OUT selection output

Non-voltage input × 4 inputs

Output terminal

BCD output

NPN open-collector output × 4 ports*4

Strobe output

NPN open-collector output × 4 outputs

OUT selection output

Power supply

Supplied from the controller

Environmental resistance

Operating temperature range

0 to +50°C 32 to 122°F

Operating ambient humidity

20 to 85% RH (no condensation)

Weight

760 g

*1 Up to 1 unit can be connected to the controller.
*2 NPN open-collector output rating: 30 mA max. (30 V max.), residual voltage of 0.5 V max.
*3 Non-voltage input rating: ON voltage of 1 V max., OFF current of 0.6 mA max.
*4 Selectable from BCD output (29 bits, signed), binary output (25 bits, negative numbers are represented by the two's complement), and judgment output.

Head cable/Head extension cable

Model

CB-B3*1

CB-B10*1

CB-B5E*1

CB-B10E*1

CB-B20E*1

Type

Head cable

Extension cable

Cable min. bending R

22 mm 0.87"

Cable length

3 m 9.8'

10 m 32.8'

5 m 16.4'

10 m 32.8'

20 m 65.6'

Enclosure rating

IP67 (IEC60529)*2

Material

PVC

Weight

Approx 250 g

Approx 750 g

Approx 400 g

Approx 800 g

Approx 1500 g

*1 Regarding cable extension between the head and controller: Up to 2 cables can be connected with the CB-BxxE, and these cables should be kept to a total length of 30 m 98.4' or less.
*2 This value is from a case in which the sensor head has been connected. However, the controller side connector is not included.

Model

LS-9006D

LS-9030D

Measurement target

ø0.04 mm to ø6 mm ø0.001" to ø0.24"

ø0.3 mm to ø30 mm ø0.01" to ø1.18"

Smallest detectable object

0.04 mm 0.002"

0.3 mm 0.01"

Repeatability

±0.03 µm*1

±0.1 µm 0.000004"*4

Measurement accuracy

±0.5 µm 0.000020"*2

±2 µm 0.000079"*5

Sampling cycle

16,000 samples/sec.*3

Light source

InGaN green LED

Monitor camera

Not provided

Environmental resistance

Operating temperature range

0 to +50°C 32 to 122°F

Operating ambient humidity

20 to 85% RH (No condensation)

Operating ambientluminance

Incandescent lamp/fluorescent lamp 3,000 lux or lower

Vibration

10 to 55 Hz, double amplitude 1.5 mm 0.06", 2 hours in each direction (X,Y, and Z)

Shock resistance

15G/6 ms

Enclosure rating

IP67 (including connector part)

Material

Aluminum

Weight

Approx 4.8 kg

Approx 9 kg

*1 A ±2σ margin of error when measuring a ø1.0 mm ø0.04" rod in the center of the measurement area using outer diameter mode with the average measurement number set as 2048 times.
*2 Margin of error when a moving ø1.0 mm ø0.04" rod is measured in the 2 mm × 2 mm 0.08" × 0.08" measurement area.
*3 The sampling cycle is changed by the number of OUT set, and by the use of the mutual interference prevention function.
*4 A ±2σ margin of error when measuring a ø10 mm ø0.39" rod in the center of the measurement area using outer diameter mode with the average measurement number set as 2048 times.
*5 Margin of error when a moving ø10 mm ø0.39" rod is measured in the 10 mm × 10 mm 0.39" × 0.39" measurement area.

Head

Model

LS-9120M

Measurement range

0.8 mm to 120 mm 0.03" to 4.72"

Smallest detectable object

0.8 mm 0.03"

Transmitter/receiver distance

400 ±100 mm 15.75" ±3.937"

Repeatability

±0.3 µm 0.000012"*1

Measurement accuracy

±8 µm 0.000315"*2

Sampling cycle

16000 samples/sec.

Light source

InGaN green LED

Monitor camera

Provided

Environmental resistance

Operating temperature range

0 to +50°C 32 to 122°F

Operating ambient humidity

20 to 85% RH (no condensation)

Operating ambientluminance

Incandescent lamp/fluorescent lamp 3000 lux or lower

Vibration

10 to 55 Hz, double amplitude 1.5 mm 0.06", 2 hours in each direction (X,Y, and Z)

Shock resistance

15G/6 ms

Enclosure rating

IP67 (including connector)

Material

Aluminum

Weight

Transmitter: Approx 1800 g, Receiver: Approx 2800 g, Base: Approx 1600 g

*1 A ±2σ margin of error when measuring a ø40 mm ø1.57" rod in the center of the measurement area using outer diameter mode with the average measurement number set as 2048 times.
*2 Margin of error when a moving ø40 mm ø1.57" rod is measured in the 40 mm × 120 mm 1.57" × 4.72" measurement area using outer diameter mode.

Measurement Library