High-speed 2D Optical Micrometer

TM-3000 series

Select Language

Model

TM-006

TM-040

TM-065

Measuring range

ø6 mm 0.24"

ø40 mm 1.57"

ø65 mm 2.56"

Minimum detectable object

0.04 mm 0.001"

0.3 mm 0.01"

0.5 mm 0.02"

Transmitter/receiver distance

60 mm 2.36"

180 mm 7.09"

270 mm 10.63"

Light source

GaN Green LED

InGaN Green LED

Measurement accuracy

±0.5 µm ±0.000020" *1

±2 µm 0.000079" *5

±3 µm ±0.000119"*7

Repeatability

±0.06 µm*2

±0.15 µm ±0.000006" *6

±0.2 µm 0.000008" *8

Sampling cycle (trigger interval)

5.5ms (33ms at the initial setting)*3

Environmental resistance

Enclosure rating

IP64*4

Operating temperature range

0 to 50°C 32 to 122°F

Operating ambient humidity

35 to 85% RH (No condensation)

Material

Aluminum

Weight

Transmitter

Approx 140 g

Approx 560 g

Approx 1280 g

Receiver

Approx 340 g

Approx 720 g

Approx 1460 g

Base

Approx 220 g

Approx 630 g

Approx 1500 g

*1 In a measurement area of 2 mm 0.08"× ø4 mm ø0.16" error when measuring width of KEYENCE standard object (glass calibration scale).
*2 Value of ±2σ measuring the width of KEYENCE standard object (glass calibration scale) in the centre of the measurement area, an average 16 times, average 1.3 mm 0.05" line.
*3 When measurement area is minimum, others are initial settings
*4 Apart from connector component
*5 In a measurement area of 10 mm 0.39"× ø26 mm ø1.02" error when measuring width of KEYENCE standard object (glass calibration scale).
*6 Value of ±2σ measuring the width of KEYENCE standard object (glass calibration scale) in the center of the measurement area, an average 16 times, average 8 mm 0.31" line.
*7 Error when measuring width of KEYENCE standard object (glass calibration scale) in a measurement area of 20 mm 0.79"× ø40 mm ø1.57".
*8 Value of ±2σ measuring the width of KEYENCE standard object (glass calibration scale) in the center of the measurement area, an average 16 times, average 14 mm 0.55" line.

Model

TM-3001

TM-3001P

Head compatibility

Compatible

Number of connectable sensors

2 units max.*1

Display

Minimum display unit

0.01 μm, 0.001 mm2, 0.01°

Maximum display range

±9999.99 mm ±393.7", ±99999.9 mm2, ±99999.9°

Input terminal block

Laser remote interlock input

Non-voltage input*2

Trigger input (for Head A)

Non-voltage input*2

Voltage input*2

Timing 1 input

Auto-zero 1 input

Reset input

Output terminal block

Analog voltage output

±10 V x 2 outputs, out put impedance: 100 Ω*2

Total judgment output

NPN open-collector output*2

PNP open-collector output*2

Error output

NPN open-collector output (N.C.)*2

PNP open-collector output (N.C.)*2

Process output

NPN open-collector output*2

PNP open-collector output*2

Trigger input enable output

Adjusted error output

Expansion connector

Trigger input (for Head A)

Non-voltage input*2

Voltage input*2

Timing 2 input

Auto-zero 2 input

Program switching input

Non-voltage input, 4 inputs*2

Voltage input, 4 inputs*2

Memory card save input

Non-voltage input*2

Voltage input*2

Judgment/Binary output

3-level judgment output: OUT1 to OUT16, total judgment output
Binary output: OUT1 to OUT16 measured data output (21 bits)
NPN open-collector output*3

3-level judgment output: OUT1 to OUT16, total judgment output
Binary output: OUT1 to OUT16 measured data output (21 bits)
PNP open-collector output*3

Strobe output

NPN open-collector output*2

PNP open-collector output*2

Trigger input enable output

Analog RGB monitor output

SVGA (800 x 600 pixels)

RS-232C interface

Measured data output and control input/output (Maximum baud rate: 115200 bps, selectable)

USB interface

In conformity with USB Revision 2.0 HI-SPEED (USB 1.1 Full-SPEED compatible)

Ethernet interface

1000BASE-T/1000 BASE-TX/10 BASE-T

Memory card

SD card CA-SD4G (4GB), CA-SD1G (1GB) support

Rating

Power voltage

24 VDC ±10%, Ripple: 10% (P to P) or less

Current consumption.

1 head connected 480mA max./ 2 heads connected 550mA max.

Environmental resistance

Operating temperature range

0 to 50°C 32 to 122°F

Operating ambient humidity

35 to 85% RH (No condensation)

Material

Polycarbonate

Weight

Approx 1120 g

*1 1 or 2 units can be connected only with the same head model
*2 • The rating of the NPN/PNP open collector output (output terminal block): 50 mA (30 V or less) max., residual voltage: 1.4 V or less (50 mA) 1.0 V (20 mA)
• The rating of the NPN/PNP open collector output (expansion connector): 50 mA (30 V or less) max., residual voltage: 1.0 V or less
• Rating for non-voltage input, ON voltage 1V max., OFF current 0.3mA max. (trigger input terminal, ON voltage 5V max., OFF current 1mA max.)
• Voltage rating, maximum rating 26.4V, ON voltage 10.8V, OFF current 0.3mA (trigger input terminal maximum rating 26.4V, ON voltage 10.8V, OFF current 1mA)
*3 • The rating of the NPN/PNP open collector output (output terminal block): 50 mA (30 V or less) max., residual voltage: 1.4 V or less (50 mA) 1.0 V (20 mA)
• The rating of the NPN/PNP open collector output (expansion connector): 50 mA (30 V or less) max., residual voltage: 1.0 V or less
• Rating for non-voltage input, ON voltage 1V max., OFF current 0.3mA max. (trigger input terminal, ON voltage 5V max., OFF current 1mA max.)
• Voltage rating, maximum rating 26.4V, ON voltage 10.8V, OFF current 0.3mA (trigger input terminal maximum rating 26.4V, ON voltage 10.8V, OFF current 1mA). OUT 1 to OUT 8 decision result, OUT 9 to OUT 16 decision result, time share output of binary measurement data.