Micro-head Spectral-interference Laser Displacement MeterSI-F series

SI-F01U

Spectroscopy Unit

SPECIFICATIONS

Select Language

Model

SI-F01U

Type

Micro-head type Spectrum unit

Measurement range

0.05 to 1.1 mm 0.002 to 0.04" *1

Light source

Infrared SLD Central wavelength 820 nm Output 0.6 mW Class 1 Laser product (IEC60825-1/FDA (CDRH) Part 1040.10 *2 )

Spot diameter

ø20 µm ø0.000787"*3

Linearity

±0.2 µm ±0.000008" *4

Resolution

0.001 µm *5

Sampling cycle

200 µs

Light source for guide

Red semiconductor laser Wavelength 650 nm Output 0.1 mW Class 1 Laser product (IEC60825-1/FDA (CDRH) Part 1040.10 *2 )

LED display

Target near center of measurement range: green lights. Target within measurement range: orange lights. Target outside measurement range: Flashes orange.

Temperature fluctuation

0.01% of F.S./°C

Environmental resistance

Enclosure rating

Operating ambientluminance

Incandescent lamp or fluorescent lamp: 10000 lux max.

Operating temperature range

0 to +35°C 32°F to 95°F

Operating ambient humidity

35 to 80%RH (No condensation)

Vibration

10 to 55 Hz, 0.5 mm 0.02" double amplitude in X, Y, and Z directions, 2 hours respectively

Material

Polycarbonate

Weight

Approx. 1 kg

*1 Indicates distance from the front of the sensor head. For long distance type sensor heads there is an individual difference of up to ±0.2 mm ±0.01".
*2 The laser classification for FDA (CDRH) is implemented based on IEC60825-1 in accordance with the requirements of Laser Notice No. 50.
*3 Indicates the minimum beam spot diameter within the measurement range.
*4 This value is obtained when the measurement target is a glass plate with the number of averaging measurements set to 256.
*5 This value is obtained by measuring a glass plate surface located at the center of the measurement range with the number of averaging measurements set to 4,096.