Separate-amplifier type digital flow sensorFD-V70 series

FD-V70AP

Amplifier Unit, DIN Rail Mount Type, PNP

SPECIFICATIONS

Select Language

Model

FD-V70AP

Type

DIN-rail mounting type

Current consumption

At 12 V

Normal

1500 mW (125 mA) or less*1

Eco mode

1320 mW (110 mA) or less*1

At 24 V

Normal

2000 mW (83 mA) or less*1

Eco mode

1700 mW (70 mA) or less*1

Display power

3 digit dual row 7-segment LED (character height: Upper characters; 8 mm 0.315", red. Lower characters; 5.7 mm 0.224", green)Display refresh cycle: 5 times/second (10 times/second when response time is 0.5 seconds in the standard display mode)

Operation status indicators

Red LED × 3 (corresponding to control output 1, control output 2, and the abnormality alarm output)

Detecting area

FD-P05

0.5 to 5 L/min

FD-P20

2 to 20 L/min

FD-P50

5 to 50 L/min

FD-P100

10 to 100 L/min

FD-F04

0.4 to 4 L/min

FD-F20

2 to 20 L/min

FD-F50

5 to 50 L/min

Display resolution

Standard display mode oraveraged display mode

FD-P05

0.01

FD-P20

0.1

FD-P50

FD-P100

FD-F04

0.01

FD-F20

0.1

FD-F50

Averaged 2-digit displaymode

FD-P05

FD-P20

1

FD-P50

FD-P100

FD-F04

0.1

FD-F20

1

FD-F50

Hysteresis

Variable

Response time (chatter prevention feature)

Selectable from 0.5, 1, 2, 5, 20 and 60

Accumulated flow unit

FD-P05

0.01, 0.1, 1, 10, 100Selectable fromthe above

FD-P20

0.1, 1, 10, 100, 1000Selectable from the above

FD-P50

Integratedflow unit

FD-P100

Accumulated flow unit

FD-F04

0.01, 0.1, 1, 10, 100Selectable fromthe above

FD-F20

0.1, 1, 10, 100, 1000Selectable from the above

Integratedflow unit

FD-F50

Integration reset/bank input

Input time 20 ms or greater (Analog output/abnormality alarm output switchable)

Control output/integration pulse output

PNP open collector Max. 50 mA (30V or less) Residual voltage 1 V or less, 2 outputs (N.O./N.C. switchable)

Abnormality alarm output

Analog output

4 to 20 mA Maximum load resistance 260 ΩAnalog output range can be set with the maximum 10% step of F.S.

Repeatability

±0.5% of F.S. or less (Fluid temperature 20°C 68°F at the ambient operating temperature of 25°C 77°F)

Rating

Power voltage

12 to 24 VDC, ripple (P-P): 10% or less

Environmental resistance

Operating temperature range

0 to 50°C 32 to 122°F, no freezing

Operating ambient humidity

35 to 85% RH (No condensation)

Vibration

10 to 55Hz Compound amplitude 1.5 mm 0.06" XYZ 2 hours for each axis

Material

Main body and cover: polycarbonate, Key top: silicon rubber

Accessories

Power cable, mounting bracket, Instruction Manual

Weight

about 85 g (including power cable)

*1 Excluding analog output current.
Including head block.