Sensing Guide Pulse Level SensorFL series

FL-S001

Amplifier Unit, Sanitary Specification

SPECIFICATIONS

Select Language

Model

FL-S001

Type

Sanitary type

Measurement range

200 to 2000 mm 7.87" to 78.74"

Specific conductive capacity of measurable media

2 or more*1

Resolution

1 mm 0.04"*2

Linearity

±3 mm ±0.12"*2

Temperature characteristics

0.1 mm / °C

Undetectable area

In the case of water

Top side

25 mm 0.98"

Bottom side

45 mm 1.77" max.*3

In the case of oil

Top side

50 mm 1.97"

Bottom side

65 mm 2.56" max.*3

Judgment output response time

0.4 s minimum

Probe horizontal torque

6 N m

Tank pressure

-0.1 to + 1 Mpa

Connection port

2S ferrule

Control output

Judgment output/alarm output

NPN / PNP open collector (selectable)
30 VDC max., 50 mA max. for each
Residual voltage: 2 V max. for NPN, 2.5 V max. for PNP, N.O. / N.C. switchable

Analog output

4 to 20 mA, maximum load resistance: 350 Ω (Response time: 0.1 s after comparator output determined [90% response])
Ripple when target remains stationary (P-P) 2% of F.S. max.

Analog output accuracy

Resolution

1 mm 0.04"

Zero accuracy

±0.1 mA (Zero point = 4 mA)

Full scale accuracy

±0.2 mA (full scale = 20 mA)

Applicable cable

8-pin M12 connector

Rating

Power voltage

10 to 30 VDC, ripple (P-P) 10% included, Class 2 or LPS

Current consumption

300 mA max. (at 10 V) / 120 mA max. (at 30 V) (excluding load)

Environmental resistance

Enclosure rating

IP67

Ambient temperature

-20 to + 60 °C -4 to + 140°F

Relative humidity

35 to 85 % RH (No condensation)

Environmentalresistance

Operating medium temperature

-20 to + 150°C -4 to 302°F

Environmental resistance

Vibration resistance

10 to 55 Hz, 0.75 mm 0.03" double amplitude in X, Y and Z directions, 2 hours respectively

Shock resistance

300 m / s2 in 6 directions, 3 times respectively

Material

Liquid end materials

Sheath: PFA
Probe: SUS304
(do not connect this directly)

Case

Metal section of the housing: SUS304
Plastic section of the housing: PBT, PAR, HNBR, EPDM, PTFE, PPSU, PET

Weight

670 g

*1 These values were measured at the position of 100 mm 3.94" or less from the wall of a metal tank. Contact KEYENCE for the case of a plastic tank.
*2 These values were measured under the following conditions.
Medium: Water, Tank material: Metal, Linearity range: Between 6 cm 2.36" from the top and 1 cm 0.39" from the bottom, Average time: 4 s
*3 For the sanitary / plastic types, the undetectable area from the bottom end varies depending on the length of the probe.

  • General Catalog Request
  • Trade Shows and Exhibitions
  • eNews Subscribe

Process Controls / Process Sensors