Recent Searches
    Series ( results)
      Models ( results)
        Keywords

          3D Scanner CMMVL series

          VL-350

          3D Scanner CMM Controller

          SPECIFICATIONS

          Select Language

          Model

          VL-350

          Type

          Controller

          Measurement range

          Low magnification: ø300 mm × H 200 mm ø11.81" × H 7.87"
          High magnification: ø70 mm × H 50 mm ø2.76" × H 1.97"

          Display resolution

          0.1 µm*1

          Repeatability (σ)

          2 µm 0.000079"*2

          Measurement accuracy

          ±10 µm ±0.0004"*3

          Image sensor

          Type 1, 4-megapixel monochrome CMOS image sensor

          Optical transmitter system

          Offset transmission-type projector module

          Optical receiver system

          Low/high magnification offset dual optical module

          Light sources

          Observation light source

          LEDs in 3 colors (red, green, blue)

          Measurement light source

          Blue LED

          Zoom

          Digital - 0.1x to 8x

          Display size

          Video: 2048 × 2048 pixels; Still images: 2048 × 2048 pixels

          Stage

          Rotation mechanism

          360° (infinite rotation)

          Tilt mechanism

          Up to 45°

          Load capacity

          10 kg

          Control PC

          Dedicated PC specified by KEYENCE (OS: Windows® 10)*4

          Power supply

          Power voltage

          100 to 240 VAC ±10%, 50/60 Hz

          Power consumption

          200 VA

          Ambient temperature range

          +15 to +30°C +59 to + 86°F

          Ambient humidity range

          35 to 80% RH (assuming no condensation)

          Weight

          3.8 kg

          *1 Common for X, Y and Z axes
          *2 Value when measurement is performed using the measurement mode specified by KEYENCE using the standard gauge specified by KEYENCE and at an operating ambient temperature of 23°C ±1°C +73.4°F ±1.8°F
          *3 In reference to JIS B 7440-8, Value when measurement is performed using the measurement mode specified by KEYENCE using the standard gauge specified by KEYENCE and at an operating ambient temperature of 23°C ±1°C +73.4°F ±1.8°F
          *4 Windows® 10 Professional