          SPECIFICATIONS

          Select Language

          Model

          IM-6225

          Type

          Head

          Image sensor

          1" 6.6 mega pixel monochrome CMOS

          Display

          10. 4" LCD monitor (XGA: 1024 × 768)
          External monitor connectable (clone output)

          Receiver lens

          Double telecentric lens

          Image
          measurement

          Field of view

          Wide field: ø100 × L200 mm ø3.94" × 7.87", High precision: 25 × L125 mm 0.98" × 4.92"

          Minimum display unit

          0.1 µm

          Repetition accuracy

          W/o stage movement

          Wide field: ±1 µm, High precision: ±0. 5 µm

          With stage movement

          Wide field: ±2 µm, High precision: ±1. 5 µm

          Measurement accuracy (±2σ)

          W/o binding

          Wide field: ±5 µm*1, High precision: ±2 µm*2

          With binding

          Wide field: ±(7 + 0. 02 L) µm*3, High precision: ±(4 + 0. 02 L) µm*4

          Height measurement

          Measuring range

          -

          Measurement force

          Measurement position accuracy (XY)

          Minimum display unit

          Measurable area (XY)

          Standard probe

          Wide-range probe

          Repetition accuracy

          Standard probe

          Wide-range probe

          Measurement accuracy

          Standard probe

          Wide-range probe

          External remote input

          Non-voltage input (with and without contact)

          External output

          OK/NG/FAIL/MEAS

          Relay output/rated load: 24 VDC, 0. 5 A/ON resistance

          Interface

          LAN

          RJ-45 (10BASE-T/100BASE-TX/1000BASE-T)

          USB 2.0 series A

          6 ports (front: 2, rear: 4)

          Record

          Hard disk drive

          250 GB

          Illumination system

          Transparent

          Telecentric transparent illumination

          Ring

          Four division, multi-angle illumination (electric)
          Slit ring (directivity) illumination (electric)

          X stage

          Moving range

          100 mm 3.94" (electric)

          Z stage

          30 mm 1.18" (electric)

          Withstand load

          2 kg

          Power supply

          Power voltage

          100 to 240 VAC 50/60 Hz

          Power consumption

          310 VA max.

          Environmental resistance

          Ambient temperature

          +10 to +35 °C 50 to 95 °F

          Relative humidity

          20 to 80 % RH (No condensation)

          Weight

          Approx 31 kg

          *1 In the range of ø80 mm ø3.15" from the center of the stage within the operating ambient temperature range of +23°C ±1.0°C +73.4°F ±33.8°F at the focused focal point position
          *2 In the range of ø20 mm ø0.79" from the center of the stage within the operating ambient temperature range of +23°C ±1.0°C +73.4°F ±33.8°F at the focused focal point position
          *3 In the range of ø80 × 180 mm ø3.15" × 7.09" from the center of the stage, within the operating ambient temperature range of +23°C ±1.0°C +73.4°F ±33.8°F at the focused focal point position, and with a load weighing 1 kg or less on the stage (L = amount of stage movement in mm units)
          *4 In the range of ø20 × 120 mm ø0.79" × 4.72" from the center of the stage, within the operating ambient temperature range of +23°C ±1.0°C +73.4°F ±33.8°F at the focused focal point position, and with a load weighing 1 kg or less on the stage (L = amount of stage movement in mm units)