          High-Speed, High-Accuracy CCD Laser Displacement SensorLK-G3000 series

          LK-G407

          Sensor Head: High-speed Long Distance, Wide beam

          SPECIFICATIONS

          Select Language

          Model

          LK-G407

          Reference distance

          Diffused reflection: 400 mm 15.75", Specular reflection: 398 mm 15.67"

          Measuring range

          Diffused reflection: ±100 mm ±3.94", Specular reflection: ±99 mm ±3.90"*1*2

          Light source

          Type

          Red semiconductor laser

          Wavelength

          655 nm (visible light)

          Laser Class

          Class II (FDA CDRH 21CFR PART1040.10)

          Output

          0.95 mW

          Spot diameter (at reference distance)

          Approx. 290 x 8,300 µm 0.0114" x 0.3268"

          Linearity

          ±0.05 % of F.S. (F.S.= ±100 mm ±3.94")*3

          Repeatability

          2 µm 0.000079"*4

          Sampling frequency

          20 / 50 / 100 / 200 / 500 / 1,000 µs (Selectable from 6 levels)

          LED display

          Near the center of the measurement: Green lights
          Within the measurement area: Orange lights
          Outside the measurement area: Orange flashing

          Temperature characteristics

          0.01 % of F.S./°C (F.S.= ±100 mm ±3.94")

          Environmental resistance

          Enclosure rating

          IP67 (IEC60529)

          Ambient light

          Incandescent lamp or Fluorescent lamp: 5,000 lux max.

          Ambient temperature

          0 to +50 °C 32 to 122 °F

          Relative humidity

          35 to 85 % RH (No condensation)

          Vibration resistance

          10 to 55 Hz, Double amplitude 1.5 mm 0.06", 2 hours in each of the X, Y, and Z directions

          Material

          Aluminum die-cast

          Weight

          Approx. 380 g (including the cable)

          *1 The range is obtained by measuring KEYENCE's standard target (ceramic).
          *2 When the sampling rate is 20 µs, the range becomes -70 mm -2.76" (NEAR side) to -100 mm -3.94" (NEAR side) for Diffused reflection.
          When the sampling rate is 20 µs, the range becomes -70 mm -2.76" (NEAR side) to -99 mm -3.90" (NEAR side) for specular reflection.
          *3 The range is obtained by measuring KEYENCE's standard target (ceramic) with the Standard mode.
          *4 The range is obtained by measuring KEYENCE's standard (SUS) with 4,096 times of averaging at the reference distance.