          Model

          LT-9501H

          LT-9001H

          LT-9501HSO(5654)

          LT-9001HSO(5655)

          Measuring unit compatibility

          Measuring units are interchangeable without factory recalibration.

          Display

          Minimumdisplayable unit

          0.1 µm *1, 1 µm2, 0.01°

          0.1 µm *1

          Display range

          ±9999.99 µm, ±999999 µm2, 9999.99°

          ±9999.99 µm

          Microscopefunction

          Available

          Available

          Display cycle

          10 times/sec.*2

          Terminal block

          Analog output

          ±10 V x 2 outputs, output impedance: 100 Ω

          Timing input

          Non-voltage input

          Reset input

          Auto-zero input

          Monitor dedicated power supply

          24 VDC*3

          Control I/O

          Judgment mode

          3-step limits output

          For OUT1 and OUT2, and NPN open collector output*4

          Binary mode

          Binary output

          Measured data output (21 bits), OUT1/OUT2/PROFILE selectable
          NPN open collector output*4

          Strobe output

          NPN open collector output*4

          Binary selector output

          Binary selection input

          Non-voltage input*4

          Stability output

          NPN open collector output

          Laser remote input

          Non-voltage input

          Program switching input

          Non-voltage input x 3 inputs

          RS-232C interface

          Measured data output and control I/O (Selectable up to baud rate 115200 bps.)

          Video output

          NTSC compliant (PIN connector)

          Main functions

          Displacement mode

          Distance measurement, Transparent object thicknessmeasurement, Angle measurement, Relative anglemeasurement, Surface selection, Dark-out, Mask, Trend graphdisplay, and Scan width/interval change*5

          Distance measurement, Transparent object thickness
          measurement, Surface selection, Dark-out, Mask, Trend graph
          display, and Scan width/interval change*6

          Profile mode

          Area selection (Average, Maximum, Minimum, Maximum-tominimum,
          Area) Area calculation, Scan width/interval change,
          Dark-out, Smoothing, Averaging, and Profile data output*6

          *6

          Common

          Light intensity accumulation, Microscope (LT-9501H, LT-9501HSO (5654) only), Tolerance
          judgment, 8-program registration, Calibration, Averaging, Hold modes, Auto-zero, and
          interface language selection

          Rating

          Power voltage

          100 to 240 VAC ±10 %, 50/60 Hz

          Power consumption

          110VA or less

          Environmental resistance

          Pollution degree

          2

          Overvoltage category

          II

          Ambient temperature

          0 to +35 °C 32 to 95 °F

          Relative humidity

          35 to 85 % RH (No condensation)

          Weight

          Approx. 2.4 kg

          *1 Display changes every ±0.3 µm.
          *2 Varies depending on the setting
          *3 Dedicated power supply for the monitor specified by KEYENCE.
          *4 Select either the Limits mode or the Binary mode.
          *5 Select either the Distance mode or the Profile mode. (Distance mode is only available with the LT-9501HSO (5654)and LT-9001HSO (5655).)The rating of the NPN open-collector output is 30 mA (30 V or lower) maximum, and residual voltage is 0.5 V.The rating of the Non-voltage input is ON voltage 1 V or lower, and OFF current 0.6 mA or lower.
          *6 Select either the Distance mode or the Profile mode. (Distance mode is only available with the LT-9501HSO (5654)
          and LT-9001HSO (5655).)
          The rating of the NPN open-collector output is 30 mA (30 V or lower) maximum, and residual voltage is 0.5 V.
          The rating of the Non-voltage input is ON voltage 1 V or lower, and OFF current 0.6 mA or lower.

          Model

          LT-9011M

          LT-9011

          LT-9031M

          LT-9031

          Type

          High-accuracy

          Long-range

          Measurement range

          ±0.3 mm ±0.01"

          ±1.0 mm ±0.04"

          Reference distance

          6 mm 0.24"

          30 mm 1.18"

          Light source

          Type

          Red semiconductor laser

          Wavelength

          655 nm

          Output

          170 µW (IEC 60825-1) /3.0 µW (FDA (CDRH) Part 1040.10)

          Laser Class

          Class IIa Laser Product (FDA (CDRH) Part 1040.10)
          Class 1 Laser Product (IEC 60825-1)

          Spot diameter

          Approx. ø2 µm ø0.000079"

          Approx. ø7 µm ø0.000276"

          Scan width/interval

          0 to 1100 µm 0 to 0.0433" (6 steps)/
          1 to 10 µm 0.000039 to 0.000394" (4 steps)

          0 to 560 µm 0 to 0.022" (6 steps)/
          1 to 8 µm 0.000039 to 0.000315" (4 steps)

          Resolution

          0.3 µm 0.000012"*1

          Linearity

          ±0.5% of F.S.*1

          ±0.3% of F.S.*1

          Sampling cycle

          640 µs to 356 ms (14 steps)*2

          640 µs to 187 ms (14 steps)*2

          Display

          Minimumdisplayable unit

          0.1 µm *3, 1 µm2, 0.01°

          0.1 µm *3

          Display range

          ±9999.99 µm, ±999999 µm2, 9999.99°

          ±9999.99 µm

          Microscopefunction

          Available

          Available

          Display cycle

          10 times/sec.*4

          Terminal block

          Analog output

          ±10 V x 2 outputs, output impedance: 100 Ω

          Timing input

          Non-voltage input

          Reset input

          Auto-zero input

          Monitor dedicated power supply

          24 VDC*5

          Control I/O

          Judgment mode

          3-step limits output

          For OUT1 and OUT2, and NPN open collector output*6

          Binary mode

          Binary output

          Measured data output (21 bits), OUT1/OUT2/PROFILE selectable
          NPN open collector output*6

          Strobe output

          NPN open collector output*6

          Binary selector output

          Binary selection input

          Non-voltage input*6

          Stability output

          NPN open collector output

          Laser remote input

          Non-voltage input

          Program switching input

          Non-voltage input x 3 inputs

          RS-232C interface

          Measured data output and control I/O (Selectable up to baud rate 115200 bps.)

          Video output

          NTSC compliant (PIN connector)

          Microscopefunction

          Function presence/absence

          Kompatibel

          N/A

          Kompatibel

          N/A

          Microscope function

          Field of view

          1.3 mm x 1.05 mm0.05" x 0.04"

          2.5 mm x 2.0 mm 0.10" x 0.08"

          Type

          Available

          Unavailable

          Available

          Unavailable

          Illumination light source

          Infrared LED
          (wavelength: 870 nm)

          Infrared LED
          (wavelength: 870 nm)

          Main functions

          Displacement mode

          Distance measurement, Transparent object thicknessmeasurement, Angle measurement, Relative anglemeasurement, Surface selection, Dark-out, Mask, Trend graphdisplay, and Scan width/interval change*7

          Distance measurement, Transparent object thickness
          measurement, Surface selection, Dark-out, Mask, Trend graph
          display, and Scan width/interval change*8

          Profile mode

          Area selection (Average, Maximum, Minimum, Maximum-tominimum,
          Area) Area calculation, Scan width/interval change,
          Dark-out, Smoothing, Averaging, and Profile data output*8

          *8

          Common

          Light intensity accumulation, Microscope (LT-9501H, LT-9501HSO (5654) only), Tolerance
          judgment, 8-program registration, Calibration, Averaging, Hold modes, Auto-zero, and
          interface language selection

          Temperature characteristics (20 to 30°C (68 to 86°F))

          ±0.5% of F.S.

          ±0.25% of F.S.

          Rating

          Power voltage

          100 to 240 VAC ±10 %, 50/60 Hz

          Power consumption

          110VA or less

          Environmental resistance

          Pollution degree

          2

          Overvoltage category

          II

          Ambient light

          Incandescent lamp/Fluorescent lamp: 10,000 lux max.

          Ambient temperature

          0 to +35 °C 32 to 95 °F

          Relative humidity

          35 to 85 % RH (No condensation)

          Weight

          Approx. 400 g

          Approx. 500 g

          *1 The value when the measurement target is a mirrored surface object measured in displacement mode, scan width/interval 120 µm 0.004725"/2 µm 0.000079", and 8-times average
          *2 Sampling cycle differs according to the manufacturing variation of individual measuring units. The value when the FINE mode is set to OFF.
          *3 Display changes every ±0.3 µm.
          *4 Varies depending on the setting
          *5 Dedicated power supply for the monitor specified by KEYENCE.
          *6 Select either the Limits mode or the Binary mode.
          *7 Select either the Distance mode or the Profile mode. (Distance mode is only available with the LT-9501HSO (5654)and LT-9001HSO (5655).)The rating of the NPN open-collector output is 30 mA (30 V or lower) maximum, and residual voltage is 0.5 V.The rating of the Non-voltage input is ON voltage 1 V or lower, and OFF current 0.6 mA or lower.
          *8 Select either the Distance mode or the Profile mode. (Distance mode is only available with the LT-9501HSO (5654)
          and LT-9001HSO (5655).)
          The rating of the NPN open-collector output is 30 mA (30 V or lower) maximum, and residual voltage is 0.5 V.
          The rating of the Non-voltage input is ON voltage 1 V or lower, and OFF current 0.6 mA or lower.

          Model

          LT-C2*1

          LT-C10*1

          Cable length

          2 m 6.56'

          10 m 32.81'

          Weight

          Approx. 200 g

          Approx. 700 g

          *1 Up to 3 cables can be connected with a total maximum length of 20 m 65.61'.