Recent Searches
    Series ( results)
      Models ( results)
        Keywords

          CCD Laser MicrometerVG series

          VG-301

          Amplifier Unit

          [Discontinued model]

          This model has been discontinued.

          Recommended Replaceable Products
          To contact us:

          SPECIFICATIONS

          Select Language

          Model

          VG-301

          Light source

          Type

          Visible red semiconductor laser

          Maximum output

          38 µW

          Wavelength

          670 nm

          FDA (CDRH) 21CFR Part
          1040.10

          Class II Laser Product, Output : 38 µW

          IEC 60825-1

          Class 1 Laser Product, Output : 0.2 mW

          Measuring area

          35 mm 1.38"

          Setting distance

          0 to 300 mm 11.81"

          Linearity

          ±0.1% of F.S. (F.S. = 30 mm 1.18") *1*2

          Minimum displayable unit

          5 µm 0.20 Mil

          Minimum detectable object

          0.5 mm 0.02" dia. opaque material*3

          Repeatability

          5 µm*4*5

          Interface

          RS-232C

          Monitor output

          Output voltage

          ±7 V

          Impedance

          100 Ω

          Response time

          3.9 ms (number of averaging measurements : 1)

          Control output

          Control

          HI, GO, LO : NPN 100 mA, 40 V max.*6

          Control output

          NPN open collector : 100 mA max., (40 V max.) (one-shot output time : 1.3 ms)

          Control output

          Response time

          3.9 ms (number of averaging measurements : 1)

          Analog output

          Resolution (ripple)

          2 mV (number of averaging measurements : 4096)

          Resolution

          5 µm 0.20 Mil (number of averaging measurements : 16)
          Light-receiving element/CCD 5,000 bit, Scan time : 1.3 ms*2

          Self-timing

          NPN open collector : 100 mA max., (40 V max.) (one-shot output time : 1.3 ms)*6

          Stability output

          NPN open collector : 100 mA max., (40 V max.) (N.C.)*6

          Response time

          3.9 ms (number of averaging measurements : 1)

          Digital I/O

          RS-232C

          Key input

          10 keys

          Control input

          - Upper/ Lower limit select input
          - Auto-zero input
          - Auto-gain input
          - Reset input
          - Laser remote input voltage contact signals
          - Hold input

          Number of CCD pixels

          Scanning time

          Temperature
          fluctuation

          0.005% of F.S./°C (F.S. = 30 mm 1.18")

          Rating

          Power voltage

          24 VDC ±10 %, Ripple (P-P) 10 % or less

          Current consumption

          330 mA or less

          Environmental resistance

          Ambient light

          Incandescent lamp: 1,000 lux or less, Fluorescent lamp: 8,000 lux or less*7

          Ambient temperature

          0 to +50 °C 32 to 122 °F

          Relative humidity

          35 to 85 % RH (No condensation)

          Vibration resistance

          10 to 55 Hz, Double amplitude 1.5 mm 0.06", 2 hours in each of the X, Y, and Z directions

          Material

          Polycarbonate

          Weight

          Approx. 800 g (including cable)

          *1 F.S. (= 30 mm 1.18") indicates the range resulting from 2.5 mm 0.10" subtraction from each end of the 35 mm 1.38" measuring range.
          *2 Typical data obtained when the T-to-R distance was 100 mm 3.94", the receiver-to-target distance was 50 mm 1.97" and a knife edge was used to
          interrupt the laser beam.
          *3 Typical data obtained when the T-to-R distance was 100 mm 3.94" and the receiver-to-target distance was 50 mm 1.97".
          *4 When the number of averaging measurements is 16
          *5 This is a typical example when the distance between the transmitter and receiver has been set to 100 mm and a knife edge has been used to block light at a position that is 50 mm from the receiver. F.S. = 30 mm displays the measurement accuracy range.
          *6 NPN output can easily be converted into PNP output by connecting the optional OP-5148 PNP output converter.
          *7 Typical data obtained when the T-to-R distance was 100 mm 3.94", the receiver-to-target distance was 50 mm 1.97" and a knife edge was used to interrupt the laser beam.