Specs Air Quality Multi-Sensor MP-F series

Air Quality Multi-Sensor

Model

MP-FR10

MP-FR10R

MP-FR20

MP-FR20R

MP-FR80

Image

Type

Single model

Standard model

Single model

Standard model

Single model

Connection bore diameter

Rc3/8" (10 A)

Rc1/2" (15A)

Rc1" (25A)

Supported gases

Air and non-corrosive gases

Supported gas temperature

-5 to +50°C, 23 to 122°F

Operating pressure range

0.2 to 1 MPa, 2 to 10 bar, 29 to 145 psi *1

Pressure resistance

1.5 MPa, 15 bar, 217.5 psi

Flow rate

Detection principle

Heat based (mass flow rate)

Rated flow range (normal)

2.0 to 1000.0 L/min
0.12 to 60.00 m3/h 4.24 to 2118.88 ft3/h 0.07 to 35.31 CFM

4 to 2000 L/min
0.2 to 120.0 m3/h 7.06 to 4237.76 ft3/h 0.1 to 70.6 CFM

12 to 8000 L/min
0.7 to 480.0 m3/h 24.72 to 16951.04 ft3/h 0.4 to 282.5 CFM

Zero cut flow rate

2.0 L/min, 0.12 m3/h 4.24 ft3/h, 0.07 CFM

4 L/min, 0.2 m3/h 7.06 ft3/h, 0.1 CFM

12 L/min, 0.7 m3/h 24.72 ft3/h, 0.4 CFM

Measurement accuracy

±(1.5% of RD + 0.5% of F.S.) *2 *3

Repeatability

±1.0% of F.S. (with an averaging time of 1.0 s)

Display resolution

0.1 L/min, 0.01 m3/h 0.35 ft3/h, 0.01 CFM

1 L/min, 0.1 m3/h 3.53 ft3/h, 0.1 CFM

Response time

150 ms (63% response) *4

Averaging time

OFF/100 ms/200 ms/500 ms/1.0 s/3.0 s/5.0 s/10 s/30 s (selectable)

Pressure (Digital value, gauge pressure)

Rated pressure range

0.000 to 1.000 MPa, 0.00 to 10.00bar, 0.0 to 145.0 psi

Measurement accuracy

±2.0% of F.S.

Repeatability

±0.2% of F.S. (with a response time of 50 ms)

Display resolution

0.001 MPa, 0.01 bar, 0.1 psi

Response time

Selectable from 10 ms, 50 ms, 100 ms, 500 ms, 1.0 s, and 5.0 s *4

Humidity (Dew point)

Rated relative humidity range/ dew point range

15 to 100%RH (no condensation)/Corresponds to the relative humidity range *5 *6

Measurement accuracy

±4°C, ±7°F *5 *6

Display resolution

5%RH/1°C, 1°F *5 *6

Temperature

Measurement accuracy

±2.0°C, ±3.6°F *7

Display resolution

0.1°C, 0.1°F *7

Shut-off valve

Response time

1 s or less (closed → open)

Leakage amount

50 mL/min 0.002 CFM (N) or less

Filter/regulator (Standard model only)

Pressure adjustment range

0.15 to 0.90 MPa

0.15 to 0.90 MPa

Filtration degree

5 μm 0.196850 Mil

5 μm 0.196850 Mil

Drainage cup capacity

60 cc

100 cc

Check valve

Available

Display

Color LCD, status indicator

Data accumulation

Accumulation period

Accumulated data: approx. 2 years/Instantaneous data: approx. 2 weeks

Data reading

USB 2.0/Ethernet

I/O

Control output (Ch.1/2/3/4/5)

NPN/PNP setting switchable, open collector output 30 VDC or less, N.O./N.C. setting switchable, max. 100 mA/Ch., residual voltage: 2.5 V or less

Analog output (Ch.1/2)

4 to 20 mA, load resistance: 260 Ω or less *4

External input (Ch.2/3/6)

Short-circuit current: 1.5 mA or less, input time: 20 ms or more

Protection circuit

Power supply reverse connection protection, power supply surge protection, reverse connection protection for each I/O, surge protection for each I/O, overcurrent protection for each output

Power supply

Power voltage

24 VDC +25%/-20% (including ripple), Class 2 or LPS

Power consumption

1.2 A (300 ms) when opening/closing valve, 100 mA at all other times *8 (standalone usage; excluding load current)

Communication interface

USB2.0

Network compatibility

When using only this unit: IO-Link (Specification v1.1/COM3) *9
When connecting MP-FEN1 : EtherNet/IP™, PROFINET, Modbus/TCP, MC protocol/SLMP

Environmental resistance

Enclosure rating

IP67 (IEC60529) *10

Ambient temperature

−5°C to +50°C 23°F to 122°F (no freezing) *11

Relative humidity

35%RH to 85%RH (no condensation)

Vibration resistance

10 to 500 Hz; power spectral density: 0.204 G2/Hz; X, Y, and Z directions

Shock resistance

300 m/s2, XYZ axes, 10 times for each axis

Material

Sensor part : PET/PPS/PBT/POM/aluminum/SPHC
Filter/regulator : Aluminum/POM/HDPE/NBR/Nylon *12

Weight

Approx. 550 g 19.40 oz

Approx. 1050 g 37.04 oz (excluding drainage cup)

Approx. 1020 g 35.98 oz

Approx. 1900 g 67.02 oz (excluding drainage cup)

Approx. 1920 g 67.73 oz

*1 An upstream side pressure of 0.2 MPa or less (a filter/regulator adjustment pressure of 0.2 MPa or less for the standard model) worsens the flow rate characteristic of the shut-off valve.
*2 This value is guaranteed by KEYENCE inspection facilities while testing with air. Errors will be introduced by factors such as the gas type, gas temperature, and ambient temperature.
*3 Value in an environment with a constant temperature of 25°C 77°F and compressed air evaluated as purity class 1*1 (ISO 8573-1 (2010)/JIS B 8392-1 (2012)).±(5.5% of RD + 0.5% of F.S.) when equivalent to purity class 3*4.
*4 15 ms is added to the analog output response time.
*5 Measurable when at least 2% of F.S. worth of gas is flowing.
*6 The prescribed accuracy may not be met or the unit may be damaged if the compressed air contains a large amount of oil mist, organic solvent, or other gas. (For details, see “Gas Being Measured” on page 4.)
*7 When the flow is 5% of F.S. or higher. Errors due to heat generated by the main unit occur when the flow is too small.
*8 When connecting MP-FEA1 and/or MP-FEN1, add each device’s current consumption (2.1 A or less including load current).
*9 Use an IO-Link master that can supply 1.2 A or more. (1.6 A or less when connecting an MP-FEA1 and an MP-FEN1.) IO-Link is a registered trademark or trademark of PROFIBUS Nutzerorganisation e.V. (PNO).
*10 When protection is provided to prevent dust and liquid from intruding through the exhaust port. The IP67 enclosure rating is lost when a USB connection is established.
*11 If this product is used with a power supply voltage of 26.4 V or more, the upper limit on the ambient temperature is 45°C 113°F.
*12 Nylon is only used on the MP-Fx20R.

Back to top

Air Quality Multi-Sensor

Model

MP-FG10

MP-FG10R

MP-FG20

MP-FG20R

MP-FG80

Image

Type

Single model

Standard model

Single model

Standard model

Single model

Connection bore diameter

G3/8" (10 A)

G1/2" (15A)

G1" (25A)

Supported gases

Air and non-corrosive gases

Supported gas temperature

-5 to +50°C, 23 to 122°F

Operating pressure range

0.2 to 1 MPa, 2 to 10 bar, 29 to 145 psi *1

Pressure resistance

1.5 MPa, 15 bar, 217.5 psi

Flow rate

Detection principle

Heat based (mass flow rate)

Rated flow range (normal)

2.0 to 1000.0 L/min
0.12 to 60.00 m3/h 4.24 to 2118.88 ft3/h 0.07 to 35.31 CFM

4 to 2000 L/min
0.2 to 120.0 m3/h 7.06 to 4237.76 ft3/h 0.1 to 70.6 CFM

12 to 8000 L/min
0.7 to 480.0 m3/h 24.72 to 16951.04 ft3/h 0.4 to 282.5 CFM

Zero cut flow rate

2.0 L/min, 0.12 m3/h 4.24 ft3/h, 0.07 CFM

4 L/min, 0.2 m3/h 7.06 ft3/h, 0.1 CFM

12 L/min, 0.7 m3/h 24.72 ft3/h, 0.4 CFM

Measurement accuracy

±(1.5% of RD + 0.5% of F.S.) *2 *3

Repeatability

±1.0% of F.S. (with an averaging time of 1.0 s)

Display resolution

0.1 L/min, 0.01 m3/h 0.35 ft3/h, 0.01 CFM

1 L/min, 0.1 m3/h 3.53 ft3/h, 0.1 CFM

Response time

150 ms (63% response) *4

Averaging time

OFF/100 ms/200 ms/500 ms/1.0 s/3.0 s/5.0 s/10 s/30 s (selectable)

Pressure (Digital value, gauge pressure)

Rated pressure range

0.000 to 1.000 MPa, 0.00 to 10.00bar, 0.0 to 145.0 psi

Measurement accuracy

±2.0% of F.S.

Repeatability

±0.2% of F.S. (with a response time of 50 ms)

Display resolution

0.001 MPa, 0.01 bar, 0.1 psi

Response time

Selectable from 10 ms, 50 ms, 100 ms, 500 ms, 1.0 s, and 5.0 s *4

Humidity (Dew point)

Rated relative humidity range/ dew point range

15 to 100%RH (no condensation)/Corresponds to the relative humidity range *5 *6

Measurement accuracy

±4°C, ±7°F *5 *6

Display resolution

5%RH/1°C, 1°F *5 *6

Temperature

Measurement accuracy

±2.0°C, ±3.6°F *7

Display resolution

0.1°C, 0.1°F *7

Shut-off valve

Response time

1 s or less (closed → open)

Leakage amount

50 mL/min 0.002 CFM (N) or less

Filter/regulator (Standard model only)

Pressure adjustment range

0.15 to 0.90 MPa

0.15 to 0.90 MPa

Filtration degree

5 μm 0.196850 Mil

5 μm 0.196850 Mil

Drainage cup capacity

60 cc

100 cc

Check valve

Available

Display

Color LCD, status indicator

Data accumulation

Accumulation period

Accumulated data: approx. 2 years/Instantaneous data: approx. 2 weeks

Data reading

USB 2.0/Ethernet

I/O

Control output (Ch.1/2/3/4/5)

NPN/PNP setting switchable, open collector output 30 VDC or less, N.O./N.C. setting switchable, max. 100 mA/Ch., residual voltage: 2.5 V or less

Analog output (Ch.1/2)

4 to 20 mA, load resistance: 260 Ω or less *4

External input (Ch.2/3/6)

Short-circuit current: 1.5 mA or less, input time: 20 ms or more

Protection circuit

Power supply reverse connection protection, power supply surge protection, reverse connection protection for each I/O, surge protection for each I/O, overcurrent protection for each output

Power supply

Power voltage

24 VDC +25%/-20% (including ripple), Class 2 or LPS

Power consumption

1.2 A (300 ms) when opening/closing valve, 100 mA at all other times *8 (standalone usage; excluding load current)

Communication interface

USB2.0

Network compatibility

When using only this unit: IO-Link (Specification v1.1/COM3) *9
When connecting MP-FEN1 : EtherNet/IP™, PROFINET, Modbus/TCP, MC protocol/SLMP

Environmental resistance

Enclosure rating

IP67 (IEC60529) *10

Ambient temperature

−5°C to +50°C 23°F to 122°F (no freezing) *11

Relative humidity

35%RH to 85%RH (no condensation)

Vibration resistance

10 to 500 Hz; power spectral density: 0.204 G2/Hz; X, Y, and Z directions

Shock resistance

300 m/s2, XYZ axes, 10 times for each axis

Material

Sensor part : PET/PPS/PBT/POM/aluminum/SPHC
Filter/regulator : Aluminum/POM/HDPE/NBR/Nylon *12

Weight

Approx. 550 g 19.40 oz

Approx. 1050 g 37.04 oz (excluding drainage cup)

Approx. 1020 g 35.98 oz

Approx. 1900 g 67.02 oz (excluding drainage cup)

Approx. 1920 g 67.73 oz

*1 An upstream side pressure of 0.2 MPa or less (a filter/regulator adjustment pressure of 0.2 MPa or less for the standard model) worsens the flow rate characteristic of the shut-off valve.
*2 This value is guaranteed by KEYENCE inspection facilities while testing with air. Errors will be introduced by factors such as the gas type, gas temperature, and ambient temperature.
*3 Value in an environment with a constant temperature of 25°C 77°F and compressed air evaluated as purity class 1*1 (ISO 8573-1 (2010)/JIS B 8392-1 (2012)).±(5.5% of RD + 0.5% of F.S.) when equivalent to purity class 3*4.
*4 15 ms is added to the analog output response time.
*5 Measurable when at least 2% of F.S. worth of gas is flowing.
*6 The prescribed accuracy may not be met or the unit may be damaged if the compressed air contains a large amount of oil mist, organic solvent, or other gas. (For details, see “Gas Being Measured” on page 4.)
*7 When the flow is 5% of F.S. or higher. Errors due to heat generated by the main unit occur when the flow is too small.
*8 When connecting MP-FEA1 and/or MP-FEN1, add each device’s current consumption (2.1 A or less including load current).
*9 Use an IO-Link master that can supply 1.2 A or more. (1.6 A or less when connecting an MP-FEA1 and an MP-FEN1.) IO-Link is a registered trademark or trademark of PROFIBUS Nutzerorganisation e.V. (PNO).
*10 When protection is provided to prevent dust and liquid from intruding through the exhaust port. The IP67 enclosure rating is lost when a USB connection is established.
*11 If this product is used with a power supply voltage of 26.4 V or more, the upper limit on the ambient temperature is 45°C 113°F.
*12 Nylon is only used on the MP-Fx20R.

Back to top

Air Quality Multi-Sensor

Model

MP-FN10

MP-FN10R

MP-FN20

MP-FN20R

MP-FN80

Image

Type

Single model

Standard model

Single model

Standard model

Single model

Connection bore diameter

NPT3/8" (10 A)

NPT1/2" (15A)

NPT1" (25A)

Supported gases

Air and non-corrosive gases

Supported gas temperature

-5 to +50°C, 23 to 122°F

Operating pressure range

0.2 to 1 MPa, 2 to 10 bar, 29 to 145 psi *1

Pressure resistance

1.5 MPa, 15 bar, 217.5 psi

Flow rate

Detection principle

Heat based (mass flow rate)

Rated flow range (normal)

2.0 to 1000.0 L/min
0.12 to 60.00 m3/h 4.24 to 2118.88 ft3/h 0.07 to 35.31 CFM

4 to 2000 L/min
0.2 to 120.0 m3/h 7.06 to 4237.76 ft3/h 0.1 to 70.6 CFM

12 to 8000 L/min
0.7 to 480.0 m3/h 24.72 to 16951.04 ft3/h 0.4 to 282.5 CFM

Zero cut flow rate

2.0 L/min, 0.12 m3/h 4.24 ft3/h, 0.07 CFM

4 L/min, 0.2 m3/h 7.06 ft3/h, 0.1 CFM

12 L/min, 0.7 m3/h 24.72 ft3/h, 0.4 CFM

Measurement accuracy

±(1.5% of RD + 0.5% of F.S.) *2 *3

Repeatability

±1.0% of F.S. (with an averaging time of 1.0 s)

Display resolution

0.1 L/min, 0.01 m3/h 0.35 ft3/h, 0.01 CFM

1 L/min, 0.1 m3/h 3.53 ft3/h, 0.1 CFM

Response time

150 ms (63% response) *4

Averaging time

OFF/100 ms/200 ms/500 ms/1.0 s/3.0 s/5.0 s/10 s/30 s (selectable)

Pressure (Digital value, gauge pressure)

Rated pressure range

0.000 to 1.000 MPa, 0.00 to 10.00bar, 0.0 to 145.0 psi

Measurement accuracy

±2.0% of F.S.

Repeatability

±0.2% of F.S. (with a response time of 50 ms)

Display resolution

0.001 MPa, 0.01 bar, 0.1 psi

Response time

Selectable from 10 ms, 50 ms, 100 ms, 500 ms, 1.0 s, and 5.0 s *4

Humidity (Dew point)

Rated relative humidity range/ dew point range

15 to 100%RH (no condensation)/Corresponds to the relative humidity range *5 *6

Measurement accuracy

±4°C, ±7°F *5 *6

Display resolution

5%RH/1°C, 1°F *5 *6

Temperature

Measurement accuracy

±2.0°C, ±3.6°F *7

Display resolution

0.1°C, 0.1°F *7

Shut-off valve

Response time

1 s or less (closed → open)

Leakage amount

50 mL/min 0.002 CFM (N) or less

Filter/regulator (Standard model only)

Pressure adjustment range

0.15 to 0.90 MPa

0.15 to 0.90 MPa

Filtration degree

5 μm 0.196850 Mil

5 μm 0.196850 Mil

Drainage cup capacity

60 cc

100 cc

Check valve

Available

Display

Color LCD, status indicator

Data accumulation

Accumulation period

Accumulated data: approx. 2 years/Instantaneous data: approx. 2 weeks

Data reading

USB 2.0/Ethernet

I/O

Control output (Ch.1/2/3/4/5)

NPN/PNP setting switchable, open collector output 30 VDC or less, N.O./N.C. setting switchable, max. 100 mA/Ch., residual voltage: 2.5 V or less

Analog output (Ch.1/2)

4 to 20 mA, load resistance: 260 Ω or less *4

External input (Ch.2/3/6)

Short-circuit current: 1.5 mA or less, input time: 20 ms or more

Protection circuit

Power supply reverse connection protection, power supply surge protection, reverse connection protection for each I/O, surge protection for each I/O, overcurrent protection for each output

Power supply

Power voltage

24 VDC +25%/-20% (including ripple), Class 2 or LPS

Power consumption

1.2 A (300 ms) when opening/closing valve, 100 mA at all other times *8 (standalone usage; excluding load current)

Communication interface

USB2.0

Network compatibility

When using only this unit: IO-Link (Specification v1.1/COM3) *9
When connecting MP-FEN1 : EtherNet/IP™, PROFINET, Modbus/TCP, MC protocol/SLMP

Environmental resistance

Enclosure rating

IP67 (IEC60529) *10

Ambient temperature

−5°C to +50°C 23°F to 122°F (no freezing) *11

Relative humidity

35%RH to 85%RH (no condensation)

Vibration resistance

10 to 500 Hz; power spectral density: 0.204 G2/Hz; X, Y, and Z directions

Shock resistance

300 m/s2, XYZ axes, 10 times for each axis

Material

Sensor part : PET/PPS/PBT/POM/aluminum/SPHC
Filter/regulator : Aluminum/POM/HDPE/NBR/Nylon *12

Weight

Approx. 550 g 19.40 oz

Approx. 1050 g 37.04 oz (excluding drainage cup)

Approx. 1020 g 35.98 oz

Approx. 1900 g 67.02 oz (excluding drainage cup)

Approx. 1920 g 67.73 oz

*1 An upstream side pressure of 0.2 MPa or less (a filter/regulator adjustment pressure of 0.2 MPa or less for the standard model) worsens the flow rate characteristic of the shut-off valve.
*2 This value is guaranteed by KEYENCE inspection facilities while testing with air. Errors will be introduced by factors such as the gas type, gas temperature, and ambient temperature.
*3 Value in an environment with a constant temperature of 25°C 77°F and compressed air evaluated as purity class 1*1 (ISO 8573-1 (2010)/JIS B 8392-1 (2012)).±(5.5% of RD + 0.5% of F.S.) when equivalent to purity class 3*4.
*4 15 ms is added to the analog output response time.
*5 Measurable when at least 2% of F.S. worth of gas is flowing.
*6 The prescribed accuracy may not be met or the unit may be damaged if the compressed air contains a large amount of oil mist, organic solvent, or other gas. (For details, see “Gas Being Measured” on page 4.)
*7 When the flow is 5% of F.S. or higher. Errors due to heat generated by the main unit occur when the flow is too small.
*8 When connecting MP-FEA1 and/or MP-FEN1, add each device’s current consumption (2.1 A or less including load current).
*9 Use an IO-Link master that can supply 1.2 A or more. (1.6 A or less when connecting an MP-FEA1 and an MP-FEN1.) IO-Link is a registered trademark or trademark of PROFIBUS Nutzerorganisation e.V. (PNO).
*10 When protection is provided to prevent dust and liquid from intruding through the exhaust port. The IP67 enclosure rating is lost when a USB connection is established.
*11 If this product is used with a power supply voltage of 26.4 V or more, the upper limit on the ambient temperature is 45°C 113°F.
*12 Nylon is only used on the MP-Fx20R.

Back to top

Energy Monitor

Model

MP-FEA1

Image

Measurement specifications

Electrical service type

1-phase 2-wire, 1-phase 3-wire, 3-phase 3-wire, 3-phase 4-wire (switchable)

Number of measurement circuits

1 (up to 2 circuits with the 1-phase 2-wire setting *1)

Input specifications

Primary-side rated current, In

When using specialized CT MP-FEC100: 100 A Or, general CT (secondary-side rated input current 1 A) *2
When using specialized CT MP-FEC250: 250 A Or, general CT (secondary-side rated input current 1 A) *2
When using specialized CT MP-FEC600: 600 A Or, general CT (secondary-side rated input current 1 A) *2
When using specialized CT MP-FEC50M: 50 A
When using specialized CT MP-FEC100M: 100 A

Over current withstand capability

1.2 times the primary-side rated current (can be applied consecutively), 10 times the primary-side rated current (for 1 second or less)

Rated input voltage
(direct input)

1-phase 2-wire

・MP-FEC100/MP-FEC250/MP-FEC600, or general CT in use: 100 to 277 VAC (L-N)
・MP-FEC50M/MP-FEC100M in use: 100 to 240 VAC (L-N)

1-phase 3-wire

・MP-FEC100/MP-FEC250/MP-FEC600, or general CT in use: 100 to 240 VAC (L-N), 200 to 480 VAC (L-L)
・MP-FEC50M/MP-FEC100M in use: 100 to 240 VAC (L-N), 200 to 480 VAC (L-L)

3-phase 3-wire

・MP-FEC100/MP-FEC250/MP-FEC600, or general CT in use: 173 to 480 VAC (L-L)
・MP-FEC50M/MP-FEC100M in use: 173 to 240 VAC (L-L)

3-phase 4-wire

・MP-FEC100/MP-FEC250/MP-FEC600, or general CT in use: 100 to 277 VAC (L-N), 173 to 480 VAC (L-L)
・MP-FEC50M/MP-FEC100M in use: 100 to 240 VAC (L-N), 173 to 415 VAC (L-L)

Maximum permissible voltage to earth

・MP-FEC100/MP-FEC250/MP-FEC600, or general CT in use: 480 VAC to earth
・MP-FEC50M/MP-FEC100M in use: 240 VAC to earth

Input voltage fluctuation range

−15% to +15% of the rated input voltage

Rated input frequency

50/60 Hz

Displayable range/
Display resolution
(Main Unit)

Voltage

Display resolution 0.1 [V]

Current

Display resolution 0.001 [A]

Active power/reactive power/apparent power

-8942.400 to +8942.400 [kW/kvar/kVA], display resolution 0.001 [kW/kvar/kVA]

Active energy/reactive energy/apparent energy

0.000 to 999999.999 [kWh/kvarh/kVAh], display resolution 0.001 [kWh/kvarh/kVAh]

Power factor (PF)

-1.00 to 1.00, display resolution 0.01

Frequency

0.0 to 90.0 [Hz], display resolution 0.1 [Hz]

Measurement accuracy

Voltage

±0.5% of RD (within rated input voltage) *3

Current

・MP-FEC100/MP-FEC250/MP-FEC600, or general CT in use: ±0.5% of RD (5% ≤ In ≤ 120%), ±1.0% of RD (1% ≤ In < 5%) *2 *3
・MP-FEC50M/MP-FEC100M in use: ±0.5% of F.S. *3

Active energy

・MP-FEC100/MP-FEC250/MP-FEC600, or general CT in use: Compliant with IEC 62053-22 §7.9 table 3 class 0.5 *2 *3
Power factor 1: ±0.5% of RD (5% ≤ In ≤ 120%), ±1.0% of RD (1% ≤ In < 5%)
Power factor 0.5 inductive: ±0.6% of RD (10% ≤ In ≤ 120%), ±1.0% of RD (2% ≤ In < 10%)
Power factor 0.8 capacitive: ±0.6% of RD (10% ≤ In ≤ 120%), ±1.0% of RD (2% ≤ In < 10%)
・MP-FEC50M/MP-FEC100M in use: ±1.0% of RD (1% ≤ In ≤ 120%) *3

Reactive energy

・MP-FEC100/MP-FEC250/MP-FEC600, or general CT in use: Compliant with IEC 62053-23 §7.9 table 3 class 2 *2 *3
Reactive power factor 1: ±2.0% of RD (5% ≤ In ≤ 120%), ±2.5% of RD (1% ≤ In < 5%)
Reactive power factor 0.5 inductive or capacitive: ±2.0% of RD (10% ≤ In ≤ 120%), ±2.5% of RD (5% ≤ In < 10%)
Reactive power factor 0.25 inductive or capacitive: ±2.5% of RD (10% ≤ In ≤ 120%)
・MP-FEC50M/MP-FEC100M in use: ±2.0% of RD (5% ≤ In ≤ 120%) *3

Apparent energy

・MP-FEC100/MP-FEC250/MP-FEC600, or general CT in use: ±0.5% of RD (5% ≤ In ≤ 120%), ±1.0% of RD (1% ≤ In < 5%) *2 *3
・MP-FEC50M/MP-FEC100M in use: ±1.0% of RD (5% ≤ In ≤ 120%) *3 *4

Effect of ambient temperature

±0.03%/K *3 *5

Effect of harmonics

±0.5% of RD *3 *6

Power factor (PF)

±0.01 *3

Power

±0.1 [Hz] *3

Current consumption

36 mA

Withstand voltage

Across all terminals and housing: 2500 VAC for 1 minute
Across voltage measurement terminal and communication terminal: 2500 VAC for 1 minute
Across CT input terminal and communication terminal (When using a specialized CT): 2500 V for 1 minute

Insulation resistance

Across all terminals and housing: 20 MΩ min. (500 VDC)
Across voltage measurement terminal and CT input terminal and across communication terminal: 20 MΩ min. (500 VDC)

Environmental resistance

Ambient temperature

-5 to +55°C 23 to 131°F (no freezing)

Relative humidity

35%RH to 85%RH (no condensation)

Measurement category

II

Altitude

2500 m or less

Pollution degree

2

Mounting method

DIN rail mounting

Vibration resistance

10 to 500 Hz, power spectral concentration: 0.033G2/Hz, XYZ axes

Shock resistance

150 m/s2, 2 times each for X, Y, and Z axes

Material

Polycarbonate

Weight

Approx. 110 g 3.88 oz

*1 When measuring two circuits, use a power supply with the same voltage and phase for the two circuits. Also, use CTs of the same model.
*2 Primary-side rated current: 1 to 2000 A, secondary-side rated input current: 1 A, nominal cross-sectional area: 0.2 mm2 to 1.5 mm2 0.0003 in2 to 0.0023 in2. However, when a general CT is used, the requirements of the EU Directive(s) and CSA Certification are not met.
*3 This value assumes an ambient temperature of 23°C 73.4°F and the rated frequency. The accuracy of the CT is not included. Also, the display may be offset by an amount equal to the display resolution.
*4 Displays a guideline value when the total apparent energy is displayed with multiple energy monitors used.
*5 Excluding the time for warm up after starting communication and the temperature fluctuations attributable to measurement current fluctuations.
*6 Error when the second, third, fifth, seventh, ninth, eleventh, and thirteenth harmonics are superimposed on the base wave with current content of 30% and voltage content of 5%.

Back to top

CT (external current sensor)

Model

MP-FEC50M

MP-FEC100M

MP-FEC100

MP-FEC250

MP-FEC600

Image

Primary-side rated current (In)

50 A

100 A

250 A

600 A

Over current withstand capability

1.2 times the primary-side rated current (can be applied consecutively)

Max. connectable is correct wire diameter

ø10 0.39"

ø16 0.63"

ø24 0.94"

ø35 1.38"

Mounting method

Clamp method (split core)

Specific error on current load

±0.5% of RD (20%≤In≤100%)

±2.0% of RD (20%≤In≤100%)

±1.0% of RD (20%≤In≤100%)

Phase difference on current load

±90’ (20% In) ±90’ (100% In)

±240’ (20% In) ±180’ (100% In)

±60’ (20% In) ±60’ (100% In)

Withstand voltage

2300 VAC for 1 minute (across core and output lead)

Insulation resistance

500 VDC/100 MΩ or more (across core and output lead)

Protective element

Internal clamp element

Environmental resistance

Ambient temperature

−10°C to +55°C 14 to 131°F (no freezing)

Relative humidity

35%RH to 85%RH (no condensation)

Material

Nylon

Weight

Approx. 140 g 4.94 oz (including cable)

Approx. 170 g 6.00 oz (including cable)

Approx. 300 g 10.58 oz (including cable)

Approx. 315 g 11.11 oz (including cable)

Approx. 450 g 15.87 oz (including cable)

Back to top

Ethernet Module

Model

MP-FEN1

Image

Ethernet specifications

Compliant standard

IEEE802.3u (100BASE-TX)

Transmission rate

100 Mbps (100BASE-TX)

Transmission media

Category 5 or higher STP (shielded twisted pair)

Connector type

M12, 4-pin, female, D-code

Maximum cable length

100 m 328.1' (between nodes)

Supported networks

EtherNet/IP™, PROFINET, Modbus/TCP, MC protocol/SLMP

Number of connectible models/units

MP-F main unit: 1
MP-FEA1: up to 10
FD-R: 1

EtherNet/IP™

Supported functions

Cyclic communication (Implicit messaging)
Message communication (Explicit messaging)

Number of connections

16

Supported connections

Input Only, Exclusive Owner

RPI (communication cycle)

4 to 10,000 ms

PROFINET

Compatible network

PROFINET IO communication

Supported functions

Cyclic communication (data I/O communication)
Acyclic communication (record data communication)

Conformance class

Compliant with Conformance Class B

Conformance test version

Compliant with ver. 2.42

Applicable protocols

LLDP, SNMP, DCP

Net load

Class 2

Cyclic update time

4 to 2048 ms

MODBUS/TCP

Communication method

TCP/IP

Compatible configuration

Server

Maximum number of simultaneous client connections

4

MC protocol/ SLMP

Communication method

TCP/IP, UDP/IP

Maximum number of connections

4

Compatible frames

QnA-compatible 3E frame, 4E frame

Data format

Binary/ASCII

Power supply specifications

Connector type

M12, 4 pins, A-code, cable length: 50 cm 19.69"

Power supply voltage

24 VDC +25%/-20% (including ripple), Class 2 or LPS

Current consumption

40 mA *1

Environmental resistance

Enclosure rating

IP67 (IEC60529)

Ambient temperature

−20 to 60°C -4 to +140°F

Relative humidity

35%RH to 85%RH (no condensation)

Vibration resistance

10 to 500 Hz; power spectral density: 0.816 G2/Hz; X, Y, and Z directions

Shock resistance

100 m/s2, XYZ axes, 1000 times for each axis

Material

Case: PBT/PAR
Connector: Nickel-plated brass

Weight

Approx. 150 g 5.29 oz

*1 When connecting to the FD-R Series: 190 mA max.

Back to top